@article{jekk5866, author = {Siti Rosidah and Ari Soewondo and Mateus Adi}, title = {Optimasi Kualitas Citra Radiografi Abdomen Berdasarkan Body Mass Index dan Tegangan Tabung pada Computed Radiography}, journal = {Jurnal Epidemiologi Kesehatan Komunitas}, volume = {5}, number = {1}, year = {2020}, keywords = {Computed Radiography; Tube Voltage; Body Mass Index.}, abstract = { Background: Computed Radiography is a modality of radiographic processing that has wide-exposure latitude. So that in radiographic examination with CR modality is less concerned with the accuracy of the exposure factor especially in the use of tube voltage that can affect the image quality. Image quality determines the accuracy of diagnosis of an illness. Abdominal examination is often done in the radiological installation of Hospital, with different patient conditions especially in Body Mass Index the use of tube voltages tends to vary so as to affect the quality of radiograph and received dose Patients Methods: This type of research is analytical observational research with cross sectional research draft. Samples in research amounted to 108 distributed in each group of Body mass Index (underweight, normal, overweight) respectively 36 samples. Each body mass index group sample is taken from the use of tube voltages that are often used in RS 75 kV, 80kV, 85 kV and 90 kV respectively as much as 9. Assessments include the value of the exposure index, Noise and anatomical information. Statistical data analysis with Spearman test and univariat. Result: There is a significant connection between the use of cylinder voltage with an index exposure, noise and anatomical imagery Conclusion : Optimised tube voltage for Body Mass index Under 75 kV, normal BMI 80 kV and BMI over 85 kV }, issn = {2615-4854}, pages = {23--31} doi = {10.14710/jekk.v5i1.5866}, url = {https://ejournal2.undip.ac.id/index.php/jekk/article/view/5866} }
Refworks Citation Data :
Background: Computed Radiography is a modality of radiographic processing that has wide-exposure latitude. So that in radiographic examination with CR modality is less concerned with the accuracy of the exposure factor especially in the use of tube voltage that can affect the image quality. Image quality determines the accuracy of diagnosis of an illness. Abdominal examination is often done in the radiological installation of Hospital, with different patient conditions especially in Body Mass Index the use of tube voltages tends to vary so as to affect the quality of radiograph and received dose Patients
Methods: This type of research is analytical observational research with cross sectional research draft. Samples in research amounted to 108 distributed in each group of Body mass Index (underweight, normal, overweight) respectively 36 samples. Each body mass index group sample is taken from the use of tube voltages that are often used in RS 75 kV, 80kV, 85 kV and 90 kV respectively as much as 9. Assessments include the value of the exposure index, Noise and anatomical information. Statistical data analysis with Spearman test and univariat.
Result: There is a significant connection between the use of cylinder voltage with an index exposure, noise and anatomical imagery
Conclusion : Optimised tube voltage for Body Mass index Under 75 kV, normal BMI 80 kV and BMI over 85 kV
Article Metrics:
Last update:
The Authors submitting a manuscript do so on the understanding that if accepted for publication, copyright of the article shall be assigned to Jurnal Epidemiologi Kesehatan Komunitas (JEKK) and Magister of Epidemiology School of Postgraduate, Universitas Diponegoro as publisher of the journal. Copyright encompasses rights to reproduce and deliver the article in all form and media, including reprints, photographs, microfilms, and any other similar reproductions, as well as translations.
JEKK journal and Magister of Epidemiology School of Postgraduate, Universitas Diponegoro and the Editors make every effort to ensure that no wrong or misleading data, opinions or statements be published in the journal. In any way, the contents of the articles and advertisements published in JEKK journal are the sole responsibility of their respective authors and advertisers.
EDITORIAL OFFICE OF JURNAL EPIDEMIOLOGI KESEHATAN KOMUNITAS:
Master Program of Epidemiology, TTB A Building 5th Floor, School of Postgraduate Studies, Universitas Diponegoro, Imam Barjo Street No. 5, Semarang, Central Java, Indonesia 50241. Email : jekk.undip@gmail.com. Telp: +62-248417008
Jurnal Epidemiologi Kesehatan Komunitas (eISSN, 2615-4854) is published by Master Program of Epidemiology, School of Postgraduate Studies, Universitas Diponegoro, Semarang, Central Java, Indonesia.
View My Stats
under Creative Commons Attribution-ShareAlike 4.0 International License.