Uji Recovery Alat Rayto RT-1904C terhadap Pemeriksaan Glukosa Metode GOD-PAP

Jurusan Teknologi Laboratorium Medis, Poltekkes Kemenkes Pontianak, Pontianak, Indonesia, Indonesia

Open Access Copyright 2026 Jurnal Pengelolaan Laboratorium Pendidikan

Creative Commons License
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

Citation Format:
Abstract

Background: The accuracy of blood glucose testing is crucial in clinical diagnostics. A recovery test is a key validation parameter for evaluating the performance and reliability of an analytical instrument. This study aimed to evaluate the performance of the Rayto RT-1904C analyzer for glucose testing using the GOD-PAP method via a recovery test. Methods: This true experimental study used glucose standard samples spiked into normal and abnormal control serum at three concentrations (1%, 2%, 3%). Each concentration was measured with 9 replicates using the Rayto RT-1904C analyzer. The recovery percentage was calculated by comparing the measured value to the target value. Statistical analysis used the Mann-Whitney test. Results: The mean recovery percentages for normal control serum at 1%, 2%, and 3% concentrations were 95.81%, 98.59%, and 95.47%, respectively. For abnormal control serum, the mean values were 103.02%, 101.49%, and 97.70%. All values were within the acceptable range of 95-105%. Although the Mann-Whitney test showed a significant difference (p<0.05) between the two serum types, the instrument's performance is clinically acceptable. Conclusion: The Rayto RT-1904C analyzer demonstrated good accuracy and reliability for glucose testing with the GOD-PAP method, as evidenced by recovery values that met standard criteria.

Keywords: Glucose, GOD-PAP, Internal Quality Control, Recovery, Rayto RT-1904C

Article Metrics:

Article Info
Section: Articles
Language : ID
  1. Aini, A. N., Juwita, R., & Melani, E. (2022). Perbandingan hasil pemeriksaan glukosa darah menggunakan metode god-pap dan metode strip dilaboratorium klinik harapan sehat cianjur. Cerdika: Jurnal Ilmiah Indonesia, 2(2), 231–235
  2. Anggraini, F., Khotimah, E., & Ningrum, S. S. (2022). Analisis pemantapan mutu internal pemeriksaan glukosa darah di laboratorium RS Bhayangkara TK. I Raden Said Sukanto Tahun 2021. Binawan Student Journal, 4(1), 24–30
  3. Bintech Sinergi Medikatama. (2023). Rayto RT-1904C Semi-Automatic Chemistry Analyzer
  4. Bradshaw, A. L., Sanders, K. J., & Miles, G. A. (2023). Assessing variations in manual pipetting: An under-investigated requirement of good laboratory practice. Journal of Mass Spectrometry and Advances in the Clinical Lab, 30, 25–29. https://doi.org/10.1016/j.jmsacl.2023.09.001
  5. Ghasemi, A., & Zahediasl, S. (2012). Normality tests for statistical analysis: A guide for non-statisticians. International Journal of Endocrinology and Metabolism, 10(2), 486–489. https://doi.org/10.5812/ijem.3505
  6. Hadi, S. (2014). Validasi metode analisis: Teori dan aplikasi di laboratorium (Edisi 1). Andi
  7. Hentz, N. G., & Knaide, T. R. (2014). Effect of liquid-handling accuracy on assay performance. Journal of Laboratory Automation, 19(2), 153–162. https://doi.org/10.1177/2211068213504095
  8. Indonesia, K. K. R. (2013). Peraturan Menteri Kesehatan Republik Indonesia Nomor 43 Tahun 2013 tentang Cara Penyelenggaraan Laboratorium Klinik yang Baik. Kemenkes RI
  9. Karimi, P., Afkhami, A., & Sheikhzadeh, N. (2019). Validation of GOD PAP method for glucose determination in pathological human serum samples: Matrix effect evaluation. Journal of Clinical Laboratory Analysis, 33(5), e22845
  10. Maji, A. S., Kurniawan, M. R., & Putri, D. E. (2018). Analisis Faktor-Faktor yang Memengaruhi Pemantapan Mutu Internal pada Pemeriksaan Glukosa Darah di Laboratorium RSUD Budhi Asih. Jurnal Laboratorium Khatulistiwa, 2(1), 1–8
  11. Nurhayati, I., Riyani, A., Kurnaeni, N., Wiryanti, W., & Rinaldi, S. F. (2019). Validasi Metode GOD-PAP pada Pemeriksaan Glukosa Darah dengan Pemakaian Setengah Volume Reagen dan Sampel. Jurnal Riset Kesehatan Poltekkes Kemenkes Bandung, 11(1), 322–336. https://doi.org/10.34011/juriskesbdg.v11i1.792
  12. Putra, A. R., Nugraha, R. A., & Wibowo, Y. (2021). Validasi metode GOD PAP pada pemeriksaan glukosa darah menggunakan serum kontrol patologis: Analisis presisi, akurasi, dan recovery. Indonesian Journal of Medical Laboratory, 12(3), 152–160
  13. Rascón Careaga, A., Corella Madueño, M. A. G., Pérez Martínez, C. J., García Rojas, A. M., Souflé Vásquez, S. Z., García Moroyoqui, M. T., & García Alegría, A. M. (2021). Validation and estimation of uncertainty for a glucose determination method GOD PAP using a multi calibrator as reference. Mapan – Journal of Metrology Society of India, 36(2), 269–278. https://doi.org/10.1007/s12647-021-00441-5
  14. Shaker, G., & Zubair, M. (2025). Peroxidase-Coupled Glucose Method. StatPearls Publishing. https://www.ncbi.nlm.nih.gov/books/NBK589656/
  15. Siregar, M. T., Wulan, W. S., Setiawan, D., & Nurhati, A. (2018). Bahan Ajar Teknologi Medik (TLM) Kendali Mutu. Pusat pendidikan sumber daya manusia kesahatan
  16. Siregar, Y., Dewi, R. S., & Suriani, S. (2018). Evaluasi validasi metode pengukuran kadar glukosa darah dengan metode GOD-PAP menggunakan alat spektrofotometer. Jurnal Analis Kesehatan, 7(2), 45–51
  17. Subiyono, S., Martsiningsih, M. A., & Gabrela, D. (2016). Gambaran kadar glukosa darah metode GOD-PAP (Glucose Oxsidase-Peroxidase Aminoantypirin) Sampel Serum dan Plasma EDTA (Ethylen Diamin Terta Acetat). Jurnal Teknologi Laboratorium, 5(1), 45–48
  18. Sugiyono. (2023). Metode Penelitian Kuantitatif, Kualitatif, dan R&D (Sutopo, Ed.; Kedua). ALFABETA
  19. Suwandi, E. (2022). Metodologi Penelitian (Saputra Wijaya Andrew, Ed.; Vol. 183). PT Scifintech Andrew Wijaya
  20. WHO. (2011). Laboratory Quality Management System. WHO
  21. Widiastuti, & Riyanto, S. (2024). Perbandingan Metode Pemeriksaan Glukosa Darah Metode Heksokinase dan Peroksidase pada Tikus Wistar (Rattus norvegicus). Journal of Islamic Pharmacy, 9(1), 27–30. https://doi.org/10.18860/jip.v9i1.26890
  22. Zaidi, S. S. H., Rahim, A., Khalid, S., Ume-Farwa, & Mehmood, R. (2025). To Compare Blood Glucose Estimation Using the GOD-PAP Technique with the Hexokinase Method. Journal of Liaquat National Hospital, 3(1), 10–15

Last update:

No citation recorded.

Last update:

No citation recorded.